H2 Generator Desk

PAR series

The PAR series generators use an electrolytic cell with polymeric membrane (PEM) to produce pure hydrogen.
The innovative gas drying system requires no maintenance because these series of generators is provided with a programmable internal regeneration.
The exclusive, electronically-controlled gas/liquid separator, automatic checking for internal leaks whenever starting the unit, and constant control of operating parameters guarantee maximum safety.
Up to 20 units can be connected in parallel.
The touch-screen LCD interface provides simple and user-friendly management of all functions on the unit.

Hydrogen is produced from distilled water using a polymeric membrane (PEM). No acid or alkaline solutions are used. The drying column does not need replacing, it is regenerated automatically by activating a special cycle, lasting around 3 hours.
Regeneration needs to be performed every three to twelve months, depending on the quantity of hydrogen produced.

Product features

  • Available Flow-rates: up to 1000 cc/min
  • Outlet pressure: up to 16 bars
  • Hydrogen purity: >99.9999%
  • Drying system: Molecular sieve column with built-in regeneration system
  • Internal water tank: 2.3 litres, with electronic level control and “Autorefill” from external tank (optional)
  • Dimensions: 30x43x42(H)
  • Weight: 15 to 18.5 kg (depending on the model)
  • Certification: CE, ISO9001

Main applications

  • Carrier gas for GC and GS-MS
  • ICP-MS collision gas
  • Flame ionization detector feed gas (FID)
  • Refilling metal hydride tanks for use with fuel cells

Technical specifications

Model: PAR.H2.120 PAR.H2.180 PAR.H2.260 PAR.H2.400 PAR.H2.500 PAR.H2.1000
General data
Electrolytic cell PEM technology
H2 purity > 99.9999% ¹
Outlet pressure 12 bars (174 psi) / ( 16 bars/232 psi optional )
H2 flow-rate cc/min (Max) 120 180 260 400 500 1000
Dimensions 30x43x42(H) cm
Net weight (no water in tank) 15 kg 18.5 kg
Communication
RS232 X X X X X X
RS485 X X X X X X
USB X X X X X X
LAN Optional
Software functions
Parallel mode X X X X X X
Automatic tank refill X X X X X X
Fill canister X X X X X X
Water
Quality Deionized, ASTM II, <0.1uS
Supply pressure (Min) 0.2 bars (1.4 psi)
Supply pressure (Max) 1 bar (14 psi)
Supply flow-rate (Min, Max) 0.2 l/min, 1.5 l/min
Internal tank capacity 2.3 l
Electrical data
Type of connection IEC 320-C13
Power supply voltage 100-240 Vac 50/60 Hz
Installed power (Max) 280W 560W
Fuse rating (5×20 mm) 4A 250VAC 6.3A 250VAC
Connections
Hydrogen outlet 1/8” compression fitting
Water Quick release push-in fitting

¹ Referred to O2, dew point < -70°C. It may depends on the ambient temperature and the number of hours of work of the dryer

Information request

Need more information about this product?
Please contact us at info@microprogel.it or fill the form.